Five fields and a photograph of the product label. That is enough for a real answer. Nothing here obligates you, and we tell you whether bench work is the right call before you ship anything.
On scientific platforms, especially discontinued ones, the configuration in the laboratory is frequently not the configuration in the instrument record. Custom and OEM builds, paired multi-wavelength systems, and modules that each carry their own designation all produce records that disagree with the hardware.
What we need photographed, read individually:
A label photograph that is readable is worth more than a paragraph of description. Photograph it straight on, close, and in focus.
A power reading without its method is not usable. Before we can interpret a number we need to know whether the number is real. Tell us:
A fiber-coupled power meter used on a free-space beam collects an unknown fraction of the light. A sensor calibrated at one wavelength misreports another. A detector operating near its noise floor produces a number that looks like a measurement and is not. A measurement taken with the wrong detector, the wrong range or the wrong beam presentation can look exactly like a laser fault. Establishing which it is costs nothing at the enquiry stage and a great deal after shipping.
PhaseX Laser Services is an independent third-party service provider and is not affiliated with, authorized by, or endorsed by any manufacturer unless expressly stated in writing. Brand names and model names are used only to identify equipment for which customers may request evaluation. We do not guarantee a repair before evaluation, and we do not certify a system to original published specifications without the original acceptance data and the corresponding standards.